The JetDrive™ III drive electronics is designed to provide complex drive waveforms to MicroFab’s MJ microdispensing devices. It is computer controlled via serial communications (RS-232) and an external trigger is provided for real-time control during printing operations. An output to drive an LED strobe for drop observations is also provided, including a delay that is controlled either through the computer interface or a knob on the front panel. A Windows® based control program (JetServer™) is provided, along with the command set for customers who want to integrate control of the JetDrive™ III into their own software. An optional LabView® sample program is available. The JetDrive™ III comes in single output and multiple output configurations.
Standard Features
- Computer controlled; Windows® based control program provided, along with command set.
- 8 point bipolar trapezoidal waveform plus 12 point arb waveform mode.
- External trigger for real-time control; LED strobe output / delay.
- Integrated into VaportJet™, SphereJet™, Jetlab® II, and the Jetlab® 4 family.
Available Options
- USB relay unit for multiplexing of a single channel output.
- Multichannel units for simultaneous drive. Configurations for up to 4 and up to 16 output channels.
- LabView® sample program.
Ordering Information
CT-M3-02
JetDrive™ III controller, including command set and Windows® based Jetserver™ control program. Includes built in strobe delay. Level 02 firmware (complex waveform) included.
CT-MX-01
Eight channel relay unit for multiplexing output of JetDrive™ III. External USB model.
CT-MC3-4
Multichannel JetDrive™ III controller. Maximum number of output channels = 4.
CT-MC3-16
Multichannel JetDrive™ III controller. Maximum number of output channels =16.
JetServer™-L
LabView® based Jetserver™ sample program.
CT-SI-01
Strobe delay unit and LED. 6V power supply included.
Specifications
| |
|
| Waveform types |
8 point bipolar trapezoidal;
12 point ARB; sine |
| Pulse modes |
single; burst; continuous
|
| Voltages |
-140 to +140 V |
Rise, fall, and ARB times |
1 - 3276 µs |
| Dwell times |
3 - 3276 µs |
| Number of V,t points in ARB mode |
12 |
| Total pulse length |
< 4095 µs |
| Frequency |
1 Hz - 30 kHz |
| Resolution |
0.1V, 1µs |
| Burst count |
1-999 |
| Pulse generation control |
RS-232-C and external TTL trigger
|
| External trigger TTL |
2.5-5 V, >0.5 µs rising flank sets timing |
| Strobe delay |
-500 µs to +2500 µs relative to trigger |
| Strobe output |
1 TTL per 1-64 triggers |
| Strobe control |
programmable and manual |
| Strobe connector |
BNC |
| Trigger connector |
BNC |
| RS-232 connector |
9-pin female |
| HV connector |
DIN |
| Power |
110V or 220V (internally set) |
Brochure (PDF)
CAD drawings |
|